KARL SUSS MJB3-HP-IR
CONTACT ALIGNER
Restart Procedure
- In order, switch power off for:
- U.V. lamp power supply.
- Microscope lamp power supply.
- Contact aligner.
- Switch power on for U.V. lamp power supply. When the status panel on the unit indicates
"Ready", press the start switch.
- When the U.V. lamp light becomes visible in the lamp housing, switch on the contact aligner
power.
- Switch on the microscope lamp power supply.
- Wait ~30 minutes for lamp temperature stabilization.
- Press "CP2" on U.V. lamp power supply.